Characterization of the aerosol produced by infrared femtosecond laser ablation of polyacrylamide gels for the sensitive inductively coupled plasma mass spectrometry detection of selenoproteins

Claverie, Fanny and Pécheyran, Christophe and Mounicou, Sandra and Ballihaut, Guillaume and Fernandez Garcia, Beatriz and Alexis, Joël and Lobinski, Ryszard and Donard, Olivier F. X. Characterization of the aerosol produced by infrared femtosecond laser ablation of polyacrylamide gels for the sensitive inductively coupled plasma mass spectrometry detection of selenoproteins. (2009) Spectrochimica Acta Part B: Atomic Spectroscopy, 64 (7). 649-658. ISSN 05848547

[img]
Preview
(Document in English)

PDF (Author's version) - Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
904kB

Official URL: http://dx.doi.org/10.1016/j.sab.2009.05.027

Abstract

A 2D high repetition rate femtosecondlaserablation strategy (2-mm wide lane) previously developed for the detection of selenoproteins in gel electrophoresis by inductively coupled plasma mass spectrometry was found to increase signal sensitivity by a factor of 40 compared to conventional nanosecond ablation (0.12-mm wide lane) [G. Ballihaut, F. Claverie, C. Pécheyran, S. Mounicou, R. Grimaud and R. Lobinski, Sensitive Detection of Selenoproteins in Gel Electrophoresis by High Repetition Rate FemtosecondLaserAblation-Inductively Coupled Plasma Mass Spectrometry, Anal. Chem. 79 (2007) 6874–6880]. Such improvement couldn't be explained solely by the difference of amount of material ablated, and then, was attributed to the aerosol properties. In order to validate this hypothesis, the characterization of the aerosolproduced by nanosecond and high repetition rate femtosecondlaserablation of polyacrylamidegels was investigated. Our 2D high repetition rate femtosecondlaserablation strategy of 2-mm wide lane was found to produce aerosols of similar particle size distribution compared to nanosecond laserablation of 0.12-mm wide lane, with 38% mass of particles < 1 µm. However, at high repetition rate, when the ablated surface was reduced, the particle size distribution was shifted toward thinner particle diameter (up to 77% for a 0.12-mm wide lane at 285 µm depth). Meanwhile, scanning electron microscopy was employed to visualize the morphology of the aerosol. In the case of larger ablation, the fine particles ejected from the sample were found to form agglomerates due to higher ablation rate and then higher collision probability. Additionally, investigations of the plasma temperature changes during the ablation demonstrated that the introduction of such amount of polyacrylamidegel particles had very limited impact on the ICP source (ΔT~ 25 ± 5 K). This suggests that the cohesion forces between the thin particles composing these large aggregates were weak enough to have negligible impact on the ICPMS detection.

Item Type:Article
Additional Information:Thanks to Elsevier editor. The definitive version is available at http://www.sciencedirect.com The original PDF of the article can be found at Spectrochimica Acta website : http://dx.doi.org/10.1016/j.sab.2009.05.027
Audience (journal):International peer-reviewed journal
Uncontrolled Keywords:
Institution:Université de Toulouse > Institut National Polytechnique de Toulouse - Toulouse INP (FRANCE)
Other partners > Université de Pau et des Pays de l'Adour - UPPA (FRANCE)
Laboratory name:
Statistics:download
Deposited On:24 Jan 2013 10:05

Repository Staff Only: item control page